Product Description
- The SMC HEC-A cooling and temperature control unit is used to regulate the temperature of heat sources and process fluids.
- It is suitable for applications in laboratory, medical, analytical and manufacturing technology where a constant medium temperature is required.
- Typical applications include electron microscopes, X-ray machines, laser beam processing systems and the temperature control of chamber electrodes.
- The unit is used to cool or heat a circulating medium in order to ensure stable process conditions.
- The SMC HEC-A cooling and temperature control unit is air-cooled and, thanks to Peltier technology, enables precise control of the temperature of circulating media.
- The unit is capable of both cooling and heating and automatically switches between the two operating modes via a PID controller.
- The temperature is controlled with a high degree of stability and can also be monitored via an optional external temperature sensor.
- Depending on the model, the unit is fitted with an RS-485 or RS-232C interface for communication.
- Materials of parts in contact with the fluid – 303 stainless steel, 304 stainless steel, EPDM, ceramic, 30% glass-filled PPS, carbon, PE, polyurethane
- Cooling method – thermoelectric module (Thermo-Module)
- Radiation method – air cooling
- Control – PID controller for automatic switching between cooling and heating modes
- Ambient temperature: –10 to 35 °C
- Humidity – 35 to 80 % RH (no condensation)
- Circulating medium – pure water
- Operating temperature range: –10.0 to 60.0 °C (no condensation)
- Cooling capacity – 230 or 600 W
- Heating capacity – 600 or 900 W
- Temperature stability – ±0.01 to ±0.03 °C
- Capacity – approx. 1.2 l
- Connection size IN/OUT – Rc 1/4" or 3/8"
- Drain connection size – Rc 1/4" (with plug)
- Power supply – single-phase 100 to 240 VAC ±10%, 50/60 Hz
- Overcurrent protection – 15 A
- Current consumption – 8 A (100 V AC) to 3 A (240 V AC) or 10 A (100 V AC) to 4 A (240 V AC)
- Communication – RS-485 or RS-232C
- Accessories – mains lead, foot mounting
- Safety standards – CE marking, UL (NRTL) standards / HEC002 additionally complies with the safety standard for medical devices (IEC 60601-1)
- Length – 436 or 455 mm
- Height – 390 or 393 mm
- Width – 210 or 240 mm
- Weight – 17.5 or 27.5 kg (including base mounting)